Ultrasound Principles: Acoustic Impedance and Wave Behaviour Revision Notes | A Level AQA Physics

Acoustic impedance Z is the product of a medium's density and the speed of sound in it: Z = ρc, measured in kg m⁻² s⁻¹. Free notes and 14 questions answered.

Updated 12 Sept 2026 · Revise Right Team

Ultrasound Principles: Acoustic Impedance and Wave Behaviour is part of Non-ionising imaging (A-level only) in A Level AQA Physics.

7 min

Reading time

34

Lessons in this topic

A Level AQA

Pathway

Reviewed by Revise Right

Content review

What this lesson covers

Acoustic impedance Z is the product of a medium's density and the speed of sound in it: Z = ρc, measured in kg m⁻² s⁻¹. Free notes and 14 questions answered.

  • Define acoustic impedance and calculate it using $Z = \rho c$
  • Apply the intensity reflection coefficient equation to calculate the fraction of ultrasound reflected at a tissue boundary
  • Explain how differences in acoustic impedance between tissues determine the degree of reflection and transmission
  • Describe the role of acoustic impedance matching gel in ultrasound imaging

Key ideas to keep in view

Use this lesson to stay inside Non-ionising imaging (A-level only) while connecting the detail back to the wider Medical physics (A-level only) topic.

  • Acoustic impedance: The property of a medium that determines how much resistance it offers to the passage of a sound wave. Defined as the product of the medium's density and the speed of sound within it. Measured in kg m⁻² s⁻¹ (rayl).
  • Intensity reflection coefficient: The ratio of reflected intensity to incident intensity at a boundary between two media. Determined entirely by the acoustic impedances of the two media on either side of the boundary.
  • Attenuation: The progressive loss of ultrasound intensity as the wave travels through a medium, caused by absorption, scattering, and reflection. Greater attenuation limits the depth at which useful imaging is possible.
  • Coupling gel: A water-based gel applied between the ultrasound transducer and the patient's skin to eliminate the air gap, matching acoustic impedances and maximising transmission of ultrasound into the body.

Revision notes

Read the core explanations from Ultrasound Principles: Acoustic Impedance and Wave Behaviour before testing yourself from memory.

1. Why Boundaries Matter in Ultrasound Imaging

Picture a pulse of ultrasound travelling through the body. Every time it crosses from one tissue type into another — say, from muscle into bone — it hits a boundary. At that boundary, some of the wave bounces back towards the transducer, and some carries on deeper into the tissue. The balance between those two outcomes is what makes ultrasound imaging possible.

The key question is: what determines how much reflects and how much transmits? The answer is a property called acoustic impedance, $Z$. Understanding $Z$ is the foundation of everything else in this topic — the gel on the probe, the bright spots on the scan, and the limits of how deep ultrasound can see.

2. Acoustic Impedance: The Property That Controls Boundaries

Think of acoustic impedance as a measure of how 'resistant' a material is to being disturbed by a sound wave. A dense material through which sound travels quickly will have a high $Z$. Bone, for example, has a much higher acoustic impedance than soft tissue — and that contrast is precisely what creates a visible boundary on an ultrasound scan.

Notice that $Z$ depends on two independent properties: how dense the material is, and how fast sound moves through it. A material can have a high $Z$ because it is very dense, or because sound travels very fast within it, or both. This is why bone ($Z \approx 7.0 \times 10^6$ kg m⁻² s⁻¹) and air ($Z \approx 430$ kg m⁻² s⁻¹) sit at opposite extremes — they differ in both density and sound speed.

MediumDensity / kg m⁻³Speed of sound / m s⁻¹Acoustic impedance Z / kg m⁻² s⁻¹
Air1.33304.3 × 10²
Soft tissue (average)106015401.63 × 10⁶
Blood106015701.66 × 10⁶
Bone190036006.84 × 10⁶
Coupling gel100015001.50 × 10⁶

Scan that table and you will see something immediately useful: soft tissue, blood, and coupling gel all have very similar values of $Z$. Bone and air sit far outside that cluster. That pattern will matter enormously when we look at what happens at boundaries.

3. Reflection and Transmission at Tissue Boundaries

When an ultrasound pulse reaches a boundary between two media with acoustic impedances $Z_1$ and $Z_2$, the wave splits: part reflects back and part transmits forward. The fraction that reflects is described by the intensity reflection coefficient, $\frac{I_r}{I_i}$, where $I_r$ is the reflected intensity and $I_i$ is the incident intensity.

The equation tells a clear story: when $Z_1$ and $Z_2$ are nearly equal, the numerator $(Z_2 - Z_1)$ is close to zero, so almost nothing reflects and nearly all the intensity transmits. When $Z_1$ and $Z_2$ are very different, the numerator is large relative to the denominator, and a large fraction reflects back.

This is why the boundary between soft tissue and bone produces a strong, bright echo on a scan — the large impedance mismatch means most of the ultrasound reflects. Conversely, the boundary between blood and soft tissue produces only a faint echo, because their impedances are so close. Both outcomes are clinically useful: strong reflections locate hard structures, while weaker reflections reveal soft-tissue detail.

4. Worked Calculation: Applying the Reflection Coefficient

The reflection coefficient equation looks straightforward, but exam questions often test whether you can handle the squaring step correctly and interpret the result physically. Work through the following example carefully, paying attention to how the numbers combine.

Worked example: Calculating the Intensity Reflection Coefficient at a Tissue–Bone Boundary

An ultrasound pulse travels from soft tissue ($Z_1 = 1.63 \times 10^6$ kg m⁻² s⁻¹) into bone ($Z_2 = 6.84 \times 10^6$ kg m⁻² s⁻¹). Calculate the intensity reflection coefficient at this boundary. [3 marks]

  1. $$\frac{I_r}{I_i} = \left(\frac{Z_2 - Z_1}{Z_2 + Z_1}\right)^2$$ — [M1] Stating the correct equation earns the method mark. Always write it out in full before substituting.
  2. $Z_2 - Z_1 = 6.84 \times 10^6 - 1.63 \times 10^6 = 5.21 \times 10^6$ kg m⁻² s⁻¹ $Z_2 + Z_1 = 6.84 \times 10^6 + 1.63 \times 10^6 = 8.47 \times 10^6$ kg m⁻² s⁻¹ — Keep the powers of ten consistent. Both values share the factor $10^6$, so it cancels in the ratio — a useful check.
  3. $$\frac{5.21 \times 10^6}{8.47 \times 10^6} = \frac{5.21}{8.47} = 0.6152...$$ — The $10^6$ factors cancel, leaving a dimensionless ratio. Keep at least 4 significant figures here to avoid rounding errors in the next step.
  4. $(0.6152)^2 = 0.6152 \times 0.6152 = 0.3785...$ — [A1] Squaring is the step most often done incorrectly. Write out the multiplication explicitly: $0.6152 \times 0.6152$, not just $0.6152^2$ without checking.
  5. $$\frac{I_r}{I_i} \approx 0.38$$ — [A1] Approximately 38% of the incident intensity is reflected at this boundary. The remaining 62% transmits into the bone. This large reflection explains why bone appears as a bright, high-contrast feature on ultrasound scans.

Answer: $\frac{I_r}{I_i} \approx 0.38$ (dimensionless). About 38% of the incident intensity is reflected.

Now compare that result with what happens at a soft tissue–blood boundary. The impedances are $Z_1 = 1.63 \times 10^6$ and $Z_2 = 1.66 \times 10^6$ kg m⁻² s⁻¹. The difference is only $0.03 \times 10^6$, giving a ratio of $\frac{0.03}{3.29} \approx 0.0091$, and squaring gives $\approx 8 \times 10^{-5}$. Less than 0.01% of the intensity reflects — almost all of it transmits. That tiny echo is still detectable by modern equipment, but it is orders of magnitude weaker than the bone echo.

5. The Air Problem and the Role of Coupling Gel

Here is a practical consequence of the reflection coefficient equation that you must be able to explain. Air has an acoustic impedance of approximately $430$ kg m⁻² s⁻¹, while soft tissue has $Z \approx 1.63 \times 10^6$ kg m⁻² s⁻¹. The mismatch is enormous — a factor of roughly 3800.

Substituting into the reflection coefficient equation gives a value extremely close to 1 — meaning virtually all the ultrasound intensity reflects at an air–tissue boundary and almost none enters the body. Without intervention, the ultrasound probe pressed against dry skin would produce no useful image at all, because the thin layer of air trapped between the probe and skin would reflect essentially the entire pulse.

Coupling gel solves this problem. The gel has an acoustic impedance very close to that of soft tissue ($Z_{\text{gel}} \approx 1.50 \times 10^6$ kg m⁻² s⁻¹). Applying it between the probe and the skin eliminates the air layer, replacing the extreme air–tissue mismatch with a near-perfect impedance match. The reflection coefficient at the probe–gel and gel–skin boundaries drops to nearly zero, allowing the ultrasound pulse to enter the body with minimal reflection loss.

6. Attenuation: Why Ultrasound Cannot See Everything

Even when ultrasound enters the body successfully, its intensity decreases as it travels deeper. This progressive loss of intensity is called attenuation. Three mechanisms contribute: absorption (the medium converts wave energy into thermal energy), scattering (the wave is redirected in multiple directions by small structures), and reflection at the many internal boundaries the pulse encounters.

Attenuation increases with frequency. Higher-frequency ultrasound (typically 5–15 MHz for medical imaging) attenuates more rapidly than lower-frequency ultrasound. This creates a direct clinical trade-off: higher frequencies produce better spatial resolution because the wavelength is shorter, but they cannot penetrate as deeply before the signal becomes too weak to detect. Lower frequencies penetrate more deeply but produce coarser images.

This is why a clinician scanning a superficial structure such as a thyroid gland uses a high-frequency probe, while imaging a deep abdominal organ such as the liver requires a lower-frequency probe. The choice of frequency is always a compromise between resolution and penetration depth — both of which are governed by attenuation.

7. Exam Technique and Common Mistakes

Ultrasound questions on AQA papers follow predictable patterns. The most common calculation task is applying the reflection coefficient equation, and the most common explanation task is justifying the use of coupling gel or comparing boundaries with different impedance mismatches. Knowing where marks are awarded — and where they are lost — makes a significant difference.

A common six-mark question style asks you to compare the ultrasound imaging of two different tissue boundaries — for example, soft tissue versus bone, and soft tissue versus a fluid-filled cyst. Structure your answer by calculating or estimating the impedance mismatch at each boundary, predicting the relative reflection coefficient, and then linking this to the brightness of the echo and the clinical information obtained.

Every concept in this topic connects back to the same central idea: the greater the difference in acoustic impedance between two media, the more ultrasound reflects at their boundary. Mastering the $Z = \rho c$ and reflection coefficient equations, and being able to explain their physical consequences for imaging, covers the full scope of what AQA requires on this topic.

Frequently asked questions

How do you calculate acoustic impedance?

Acoustic impedance is Z = ρc, density times the speed of sound in the medium. Soft tissue, with a density of 1060 kg m⁻³ and a sound speed of 1540 m s⁻¹, has Z = 1060 × 1540 ≈ 1.63 × 10⁶ kg m⁻² s⁻¹.

What is acoustic impedance?

Acoustic impedance Z is the product of a medium's density and the speed of sound in it: Z = ρc, measured in kg m⁻² s⁻¹. Bone has a high value, about 6.84 × 10⁶ kg m⁻² s⁻¹, soft tissue about 1.63 × 10⁶ and air only about 430. Differences in Z between media determine how much ultrasound reflects at a boundary.

Why is gel used in an ultrasound scan?

Coupling gel eliminates the air layer between the probe and the skin. Air's acoustic impedance, about 430 kg m⁻² s⁻¹, differs from soft tissue's by a factor of roughly 3800, so virtually all the ultrasound would reflect at an air–tissue boundary. The gel's impedance, about 1.50 × 10⁶, is close to soft tissue's, so almost no reflection occurs and the pulse enters the body.

What is the acoustic impedance of air?

Air has an acoustic impedance of about 430 kg m⁻² s⁻¹, from a density of 1.3 kg m⁻³ and a sound speed of 330 m s⁻¹. The mismatch with soft tissue is a factor of roughly 3800, so an air-tissue boundary reflects almost all ultrasound.

How do you calculate the intensity reflection coefficient?

The intensity reflection coefficient, the ratio of reflected intensity to incident intensity, is ((Z₂ − Z₁)/(Z₂ + Z₁))². For soft tissue (Z = 1.63 × 10⁶) to bone (Z = 6.84 × 10⁶), the ratio is 5.21/8.47 = 0.6152, and squaring gives about 0.38: roughly 38% of the incident intensity reflects and 62% transmits into the bone.

What frequencies are used for medical ultrasound?

Medical imaging ultrasound typically uses frequencies of 5–15 MHz. Higher frequencies give better spatial resolution because the wavelength is shorter, but they attenuate more rapidly, so they cannot penetrate as deeply before the signal becomes too weak.

Why does bone show up brightly on an ultrasound scan?

Bone appears as a bright, high-contrast feature because its acoustic impedance is much larger than soft tissue's, so a large fraction, about 38%, of the incident intensity reflects at the boundary and gives a strong echo. A soft tissue–blood boundary, with nearly equal impedances, reflects less than 0.01%, giving only a faint echo.

What is ultrasound attenuation?

Attenuation is the progressive loss of ultrasound intensity as it travels deeper, caused by absorption, scattering and reflection at internal boundaries. It increases with frequency, so higher frequencies, typically 5–15 MHz, give better resolution but cannot penetrate as deeply. That is why a thyroid scan uses a high-frequency probe and a liver scan a lower-frequency one.

What are the units of acoustic impedance?

Acoustic impedance is measured in kg m⁻² s⁻¹, sometimes called the rayl. The unit comes from Z = ρc: density in kg m⁻³ multiplied by the speed of sound in m s⁻¹.

What is the acoustic impedance of bone and soft tissue?

Bone has an acoustic impedance of about 6.84 × 10⁶ kg m⁻² s⁻¹, from a density of 1900 kg m⁻³ and a sound speed of 3600 m s⁻¹. Soft tissue averages 1.63 × 10⁶ kg m⁻² s⁻¹, so the boundary between them gives a large mismatch and a strong echo.

How much ultrasound reflects at a blood and soft tissue boundary?

At a soft tissue–blood boundary, with Z values of 1.63 × 10⁶ and 1.66 × 10⁶ kg m⁻² s⁻¹, the ratio is 0.03/3.29 ≈ 0.0091, and squaring gives about 8 × 10⁻⁵. Less than 0.01% of the intensity reflects, a faint echo that modern equipment can still detect.

How much ultrasound reflects at an air and skin boundary?

At an air–tissue boundary the intensity reflection coefficient is extremely close to 1, so virtually all the ultrasound reflects and almost none enters the body. Air's impedance, about 430 kg m⁻² s⁻¹, differs from soft tissue's by a factor of roughly 3800.

When does nearly all ultrasound pass through a boundary?

Nearly all ultrasound transmits through a boundary when the two media have almost equal acoustic impedances. The numerator Z₂ − Z₁ in the reflection coefficient is then close to zero, so almost nothing reflects, as at a boundary between soft tissue and blood.

How is the ultrasound probe frequency chosen?

The probe frequency is a compromise between resolution and penetration depth. A superficial structure such as the thyroid gland is scanned with a high-frequency probe for fine detail, while a deep abdominal organ such as the liver needs a lower-frequency probe that penetrates more deeply.