MRI: Excitation, Detection, and Imaging Revision Notes | A Level AQA Physics

Gradient coils add a small varying field to B₀, so the Larmor frequency varies with position and signals can be located. Free notes and 14 questions answered.

Updated 12 Sept 2026 · Revise Right Team

MRI: Excitation, Detection, and Imaging is part of Non-ionising imaging (A-level only) in A Level AQA Physics.

8 min

Reading time

34

Lessons in this topic

A Level AQA

Pathway

Reviewed by Revise Right

Content review

What this lesson covers

Gradient coils add a small varying field to B₀, so the Larmor frequency varies with position and signals can be located. Free notes and 14 questions answered.

  • Explain how gradient field coils are used to select successive small regions of tissue during an MRI scan
  • Describe how short radiofrequency pulses cause excitation and a change of spin state in protons
  • Explain how protons emit radiofrequency signals as they de-excite and return to their original spin state
  • Describe how detected RF signals are processed by a computer to produce a visual image of a cross-section

Key ideas to keep in view

Use this lesson to stay inside Non-ionising imaging (A-level only) while connecting the detail back to the wider Medical physics (A-level only) topic.

  • Larmor frequency: The specific radiofrequency at which protons precess and absorb energy in a given magnetic field strength. Determined by the local field strength, so gradient coils make it position-dependent.
  • Gradient field coils: Electromagnets that superimpose a small, linearly varying magnetic field onto the main static field, causing the Larmor frequency to vary with position and thereby selecting a specific slice or voxel for excitation.
  • RF excitation pulse: A short burst of radiofrequency electromagnetic radiation, tuned to the local Larmor frequency, that flips proton spin states from the lower-energy alignment to a higher-energy state.
  • Relaxation (de-excitation): The process by which excited protons return to their equilibrium spin alignment, emitting RF signals that are detected and used to construct the MRI image.

Revision notes

Read the core explanations from MRI: Excitation, Detection, and Imaging before testing yourself from memory.

1. Why MRI Needs More Than a Big Magnet

A powerful static magnetic field aligns proton spins throughout the entire body — but that alone tells you nothing about where a signal comes from. To build an image, the scanner must interrogate one tiny region at a time. The clever engineering that makes this possible is the gradient field coil system, and understanding it is the key to understanding MRI imaging.

Picture the patient lying inside a uniform field $B_0$. Every proton in the body precesses at the same Larmor frequency $f_0$. If you fire an RF pulse at $f_0$, every proton in the body responds simultaneously — useless for localisation. The solution is to make the field strength vary with position, so that the resonant frequency becomes a postcode for each slice of tissue.

2. Gradient Field Coils: Giving Every Slice Its Own Frequency

Gradient field coils are sets of electromagnets wound inside the scanner bore. When energised, they add a small, linearly varying magnetic field on top of the main static field $B_0$. The total field at any position along, say, the $z$-axis becomes $B(z) = B_0 + G_z z$, where $G_z$ is the gradient strength in T m$^{-1}$.

Because the Larmor frequency is proportional to the local field strength — $f = \frac{\gamma B}{2\pi}$, where $\gamma$ is the gyromagnetic ratio of the proton — each position along the gradient now has a unique resonant frequency. A slice at position $z_1$ resonates at a slightly different frequency than a slice at $z_2$. The scanner exploits this by transmitting an RF pulse at precisely the frequency corresponding to the desired slice.

Three sets of gradient coils are used in a full scan: one for each spatial axis ($x$, $y$, $z$). By switching gradients on and off in sequence, the scanner can select a slice in one direction, then encode position within that slice in the other two directions. This is how a three-dimensional image is built up from many successive measurements of small regions.

With the gradient coils establishing a spatial map of frequencies, the scanner is ready for the next stage: actually exciting the protons in the chosen region. That requires a carefully tuned radiofrequency pulse.

3. RF Excitation: Flipping Proton Spin States

In the static field $B_0$, protons occupy one of two spin states: a lower-energy state aligned with the field (spin-up) and a higher-energy state opposed to it (spin-down). At thermal equilibrium, slightly more protons occupy the lower-energy state, producing a net magnetisation along $B_0$.

When a short RF pulse is transmitted at the Larmor frequency of the selected slice, the oscillating magnetic field of the RF wave is in resonance with the precessing protons. Energy is absorbed, and protons are promoted from the lower-energy spin state to the higher-energy spin state. This is the excitation event. The net magnetisation is tipped away from the $B_0$ direction — often by 90° for a standard pulse — into the transverse plane.

The pulse is short — typically microseconds to milliseconds — so that only the protons in the selected slice (those with the matching Larmor frequency) are significantly affected. Protons in adjacent slices, precessing at slightly different frequencies, are not in resonance and are not excited. This is the mechanism by which successive small regions are scanned: the gradient is adjusted, the resonant frequency shifts to the next slice, and a new RF pulse is applied.

Once the RF pulse ends, the excited protons do not stay in their higher-energy state. They begin to relax — and it is during this relaxation that the detectable signal is produced.

4. De-excitation: Protons Emit the Signal

After the RF pulse is switched off, the excited protons return to their equilibrium spin alignment — the lower-energy state — in a process called relaxation. As they do so, they release the energy they absorbed, emitting RF radiation at the same Larmor frequency. This emitted RF signal is the raw data of the MRI scan.

Two distinct relaxation processes occur simultaneously. Longitudinal relaxation ($T_1$) describes how the component of magnetisation along $B_0$ recovers. Transverse relaxation ($T_2$) describes how the magnetisation in the transverse plane decays as protons lose phase coherence with one another. Different tissues have characteristic $T_1$ and $T_2$ values, which is why MRI can distinguish between, for example, grey matter, white matter, and cerebrospinal fluid — even though all contain hydrogen protons.

The strength and timing of the emitted signal depend on the density of protons in the tissue and on the local chemical environment — which determines $T_1$ and $T_2$. This is why MRI provides excellent soft-tissue contrast: different tissues relax at different rates, producing signals of different amplitudes at different times after the pulse.

5. Detection and Image Reconstruction

The emitted RF signals are detected by receiver coils placed around the patient. These coils act as antennae: the oscillating magnetic field of the emitted RF radiation induces a small alternating voltage in the coil, which is then amplified and recorded. The same coil that transmits the excitation pulse often serves as the receiver — it is switched between transmit and receive modes during the scan sequence.

Because the gradient coils encode position as frequency, the detected signal from a single RF pulse contains contributions from all the protons in the selected slice, each at a slightly different frequency depending on their position within the slice. A mathematical technique called a Fourier transform is applied by the computer to decompose this complex signal into its constituent frequencies, assigning an intensity value to each spatial location.

The computer repeats this process for many different gradient settings and RF pulse timings, building up a complete data set. From this data set, it reconstructs a two-dimensional map of signal intensity across the cross-section. Higher signal intensity — reflecting more protons or favourable relaxation characteristics depending on the pulse sequence weighting used — appears brighter on the image. The result is the greyscale MRI image familiar from clinical practice.

Worked example: Tracing the MRI Signal Chain for One Slice

Describe, in sequence, the physical processes that occur from the moment the gradient coils are energised to the moment a signal is recorded for a single cross-sectional slice.

  1. $B(z) = B_0 + G_z z$ — The gradient coil superimposes a linearly varying field on $B_0$. Each position $z$ now has a unique total field strength, and therefore a unique Larmor frequency $f(z) = \frac{\gamma (B_0 + G_z z)}{2\pi}$. The desired slice at position $z_s$ has Larmor frequency $f_s$.
  2. RF pulse frequency $= f_s$ — The transmitter coil emits a short RF pulse tuned to $f_s$. Only protons in the slice at $z_s$ are in resonance; protons at other positions have different Larmor frequencies and are not significantly excited.
  3. Spin-up $\rightarrow$ spin-down (energy absorbed) — Protons in the selected slice absorb energy from the RF pulse and are promoted to the higher-energy spin state. The net magnetisation is tipped into the transverse plane. This is the excitation event.
  4. Spin-down $\rightarrow$ spin-up (energy emitted as RF) — Protons relax back to equilibrium, emitting RF radiation at $f_s$. The amplitude and decay rate of this signal depend on proton density and the $T_1$/$T_2$ relaxation times of the tissue.
  5. Induced EMF $\propto$ rate of change of flux from emitted RF — The oscillating magnetic field of the emitted RF induces an alternating voltage in the receiver coil by electromagnetic induction. This voltage is amplified and digitised.
  6. Fourier transform applied to raw signal data — The computer applies a Fourier transform to decompose the detected signal by frequency (and therefore by position within the slice). Signal intensities are mapped to spatial coordinates, producing the image of that cross-section.

Answer: Gradient coils create a position-dependent Larmor frequency → RF pulse excites the target slice → protons change spin state → protons de-excite and emit RF → receiver coil detects the signal → computer reconstructs the image.

6. Scanning Successive Regions: The Full Picture

A complete MRI scan is not a single measurement — it is a carefully choreographed sequence of gradient switching and RF pulsing, repeated for every slice and every phase-encoding step within each slice. After one slice is excited and its signal recorded, the gradient coil current is adjusted to shift the resonant frequency to the next slice position, and the process repeats. This is what the AQA specification means by 'successive small regions': the scanner works through the volume of interest systematically, one region at a time.

Within a single slice, two additional gradient directions encode the $x$ and $y$ positions of individual voxels (volume elements). The frequency-encoding gradient is applied during signal readout, spreading different columns across different frequencies. The phase-encoding gradient is applied briefly before readout, imprinting a phase shift that encodes row position. Together, these three gradient directions allow the computer to assign every detected signal to a precise three-dimensional location.

7. Exam Technique: Securing Every Mark

MRI questions at A-level frequently use the command words 'describe' and 'explain'. Examiners draw a sharp distinction between the two. A 'describe' answer states what happens in sequence — no reasons required. An 'explain' answer must include the scientific mechanism: why the gradient makes frequencies position-dependent, why only the resonant slice is excited, why de-exciting protons emit RF radiation.

Three specific language choices consistently earn marks. First, use the phrase 'change of spin state' (or 'spin flip') when describing excitation — not vague terms like 'protons are disturbed'. Second, state explicitly that the emitted RF signal has the same frequency as the excitation pulse (the Larmor frequency). Third, when describing image formation, name the computer processing step: the signals are 'processed by a computer to produce a visual image', and stronger answers mention that different tissues produce signals of different intensity due to differing proton density or relaxation times.

Frequently asked questions

What do gradient coils do in MRI?

Gradient coils add a small, linearly varying field to the main field B₀, so the total field, and therefore the Larmor frequency, varies with position. An RF pulse at one frequency then excites only the matching slice. Three sets of coils, one for each axis, select a slice and encode position within it, so successive small regions of tissue are scanned.

What are longitudinal and transverse relaxation?

Longitudinal relaxation describes how the magnetisation along B₀ recovers after the RF pulse. Transverse relaxation describes how the magnetisation in the transverse plane decays as protons lose phase coherence. Different tissues have characteristic relaxation times, which gives MRI its soft-tissue contrast.

What is phase encoding in MRI?

Phase encoding applies a gradient briefly before the signal is read out, imprinting a phase shift that encodes the row position of each voxel within a slice. Together with the slice gradient and frequency encoding, it lets the computer locate every signal in three dimensions.

What is a voxel in MRI?

A voxel is a volume element, a tiny three-dimensional region of tissue. In MRI, the three gradient directions let the computer assign every detected signal to a precise voxel, so each one gets its own signal intensity in the final image.

What is frequency encoding in MRI?

Frequency encoding applies a gradient during signal readout, so different columns within a slice emit at different frequencies. The computer then distinguishes the columns by frequency, while a phase-encoding gradient encodes row position.

How does the RF pulse excite protons in MRI?

A short RF pulse tuned to the Larmor frequency of the selected slice is in resonance with the precessing protons, so they absorb energy and change spin state from the lower-energy to the higher-energy state. This tips the net magnetisation away from B₀, often by 90°. Protons in other slices, at different Larmor frequencies, are not significantly excited.

How is the MRI signal produced and detected?

When the RF pulse is switched off, excited protons relax back to their lower-energy spin state, emitting RF radiation at the same Larmor frequency. The oscillating magnetic field of this emitted radiation induces a small alternating voltage in receiver coils around the patient by electromagnetic induction, which is then amplified and digitised.

How is an MRI image formed?

The computer applies a Fourier transform to split the detected signal into its constituent frequencies, which the gradient coils have linked to positions, and assigns an intensity to each location. Repeating this for many gradient settings builds a two-dimensional map of signal intensity across the cross-section, with stronger signals appearing brighter on the greyscale image.

Why does MRI give good soft-tissue contrast?

Different tissues have different proton densities and characteristic T₁ and T₂ relaxation times, so after the RF pulse they emit signals of different amplitudes at different times. This lets MRI distinguish between grey matter, white matter and cerebrospinal fluid, even though all of them contain hydrogen protons.

What is the flip angle in MRI?

The flip angle is the angle through which the RF excitation pulse tips the net magnetisation away from the B₀ direction, often 90° for a standard pulse, into the transverse plane. It is set by the pulse's duration and amplitude.

Why does MRI use a Fourier transform?

The signal from one RF pulse contains contributions from all the protons in the slice, each at a slightly different frequency set by its position. A Fourier transform decomposes this complex signal into its constituent frequencies, so the computer can assign an intensity value to each spatial location.

What frequency is the MRI signal?

The RF signal emitted by relaxing protons has the same frequency as the excitation pulse: the Larmor frequency of the selected slice. As protons return to the lower-energy spin state, they release the absorbed energy as radiofrequency radiation at that frequency.

How long is an MRI RF pulse?

An MRI excitation pulse is short, typically microseconds to milliseconds, so that only protons in the selected slice, with the matching Larmor frequency, are significantly affected. Protons in adjacent slices precess at slightly different frequencies and are not in resonance.

What do bright areas on an MRI image mean?

Brighter areas on an MRI image show higher signal intensity, reflecting more protons or favourable relaxation characteristics, depending on the pulse sequence weighting used. This is how grey matter, white matter and cerebrospinal fluid can be told apart, even though all contain hydrogen.