Comparing Imaging Techniques Revision Notes | A Level AQA Physics
CT and MRI give the highest spatial resolution, while ultrasound is generally considered the safest as it is non-ionising. Free notes and 15 questions answered.
Updated 12 Sept 2026 · Revise Right Team
Comparing Imaging Techniques is part of Radionuclide imaging and therapy (A-level only) in A Level AQA Physics.
12 min
Reading time
34
Lessons in this topic
A Level AQA
Pathway
Reviewed by Revise Right
Content review
What this lesson covers
CT and MRI give the highest spatial resolution, while ultrasound is generally considered the safest as it is non-ionising. Free notes and 15 questions answered.
Compare imaging techniques in terms of image resolution, convenience, and safety
Evaluate the suitability of different imaging techniques for specific clinical scenarios
Analyse the trade-offs between resolution, patient safety, and practical convenience across imaging modalities
Justify the choice of imaging technique for a given diagnostic situation using scientific reasoning
Key ideas to keep in view
Use this lesson to stay inside Radionuclide imaging and therapy (A-level only) while connecting the detail back to the wider Medical physics (A-level only) topic.
Image resolution: The ability of an imaging technique to distinguish between two closely spaced structures. Higher resolution produces finer detail in the resulting image.
Ionising radiation: Radiation with sufficient energy to remove electrons from atoms, potentially damaging DNA. Relevant to safety comparisons between imaging modalities.
Radionuclide imaging: A nuclear medicine technique in which a radioactive tracer is introduced into the body and gamma-ray emission is detected externally to produce a functional image.
Convenience (imaging context): A measure of how practical an imaging technique is, encompassing scan duration, patient preparation, equipment availability, cost, and whether the patient can remain mobile.
Revision notes
Read the core explanations from Comparing Imaging Techniques before testing yourself from memory.
1. Why Comparing Imaging Techniques Matters
Every day, clinicians face a decision: which imaging technique gives the most useful picture of what is happening inside a patient, with the least risk and the greatest practicality? Getting that decision wrong can mean a missed diagnosis, an unnecessary radiation dose, or a scan that takes so long the patient cannot tolerate it.
AQA asks you to compare imaging techniques across three specific dimensions: image resolution, convenience, and safety. These three criteria form the framework for every comparison question you will encounter, so building a precise, evidence-based understanding of each one is the goal of this lesson.
2. The Imaging Techniques You Need to Know
Before comparing techniques, you need a clear picture of what each one actually does. The main modalities covered at A-level are X-ray (including CT), MRI, ultrasound, and radionuclide imaging (including PET and gamma camera scintigraphy). Each exploits a different physical phenomenon to generate an image.
Technique
Physical basis
What is detected
X-ray / CT
Differential absorption of X-rays (ionising)
Transmitted X-ray intensity through tissue
MRI
Nuclear magnetic resonance of hydrogen nuclei (non-ionising)
Radio-wave emission from relaxing protons
Ultrasound
Reflection of high-frequency sound waves (non-ionising)
Echoes at tissue boundaries
Radionuclide / PET
Gamma emission from a radioactive tracer (ionising)
Gamma photons detected outside the body
Notice that X-ray, CT, and radionuclide imaging all involve ionising radiation, while MRI and ultrasound do not. That distinction is central to the safety comparison. Keep it in mind as you work through the next sections.
3. Image Resolution — How Much Detail Can You See?
Resolution describes the smallest separation between two structures that the technique can distinguish. A high-resolution image shows fine anatomical detail; a low-resolution image blurs nearby structures together. Resolution is not fixed — it depends on the wavelength or frequency of the probe, the detector design, and the processing algorithm used.
X-ray and CT produce high spatial resolution because X-ray wavelengths are extremely short (around $10^{-10}$ m), allowing fine structural detail to be captured. CT improves on plain X-ray by reconstructing cross-sectional slices, giving three-dimensional anatomical information with sub-millimetre resolution in modern scanners.
MRI also achieves high spatial resolution — comparable to CT for soft tissue — and has a particular advantage: it distinguishes between different soft tissues far better than X-ray does, because different tissues contain different concentrations of hydrogen nuclei and have different relaxation times. This makes MRI the preferred choice for imaging the brain, spinal cord, and joints.
Ultrasound resolution depends on the frequency of the sound wave used. Higher frequency gives shorter wavelength and therefore better resolution, but the wave is absorbed more quickly, limiting penetration depth. A 10 MHz probe resolves structures around 0.15 mm apart but cannot image deep organs effectively. A 3.5 MHz probe reaches deeper but with coarser resolution. There is always a trade-off between resolution and penetration depth in ultrasound.
Radionuclide imaging — including gamma camera scintigraphy and PET — has the lowest spatial resolution of the common modalities, typically 4–10 mm for PET and coarser still for planar scintigraphy. This is because the image is built from detected gamma photons emitted in all directions from within the body; collimators and coincidence detection improve localisation, but the fundamental limit is the statistical nature of radioactive decay and the relatively large detector elements used.
The key insight here is that radionuclide imaging sacrifices spatial resolution in exchange for something X-ray and MRI cannot provide: functional information. A PET scan shows where glucose metabolism is highest, revealing tumour activity or brain function rather than just anatomy. Resolution and function are different dimensions of image quality, and examiners expect you to acknowledge both.
4. Convenience — Practicality in a Clinical Setting
Convenience covers everything that affects how practical a scan is to perform and tolerate. AQA groups several factors under this heading: scan duration, patient preparation required, whether the patient must remain still, equipment size and availability, cost, and whether the technique can be used at the bedside or requires a specialist facility.
Plain X-ray is the most convenient modality by almost every measure. The equipment is widely available, the scan takes seconds, no special preparation is needed, and the patient can be positioned in almost any orientation. This is why X-ray remains the first-line investigation for suspected fractures and chest pathology despite its relatively limited soft-tissue contrast.
Ultrasound is also highly convenient. The equipment is portable — it can be taken to the patient's bedside or used in an outpatient clinic — and there is no ionising radiation, so it can be repeated freely. The scan is performed in real time, which is particularly useful for guiding needle biopsies or assessing foetal development during pregnancy. The main limitation is operator dependence: image quality varies significantly with the skill of the sonographer.
CT is less convenient than plain X-ray. The scanner is large, fixed in a dedicated room, and the patient must lie still inside a narrow bore for several minutes. However, a full-body CT can be completed in under 10 minutes with modern multi-slice scanners, and no special preparation is usually required beyond removing metal objects. CT is available in most district general hospitals.
MRI is the least convenient of the anatomical techniques. Scan times are long — typically 20 to 60 minutes — and the patient must remain completely still inside a loud, narrow bore magnet. Patients with certain metallic implants (pacemakers, some surgical clips) cannot safely enter the magnetic field. The equipment is expensive and available only in larger hospitals. Claustrophobic patients may require sedation.
Radionuclide imaging presents a different kind of inconvenience. The patient must be injected with a radioactive tracer and then wait for it to distribute through the body before imaging begins — this uptake period can be 30 minutes to several hours depending on the tracer. The patient is also mildly radioactive during and after the scan, which requires careful management of contact with others, particularly pregnant women and young children. PET scanners are expensive and available only in specialist centres.
5. Safety — Weighing the Risks
Safety in medical imaging centres on two main concerns: the biological effects of ionising radiation, and any non-radiation hazards specific to the technique. You need to be precise about which techniques carry which risks, and why.
Ionising radiation — used in X-ray, CT, and radionuclide imaging — can cause DNA damage. Most damage is repaired by cellular mechanisms, but a small residual risk of inducing cancer remains. The risk is proportional to the dose received, quantified in millisieverts (mSv). A chest X-ray delivers approximately 0.02 mSv; a CT of the abdomen delivers around 10 mSv; a radionuclide bone scan delivers approximately 3–4 mSv. These doses are small compared with the natural background radiation of around 2.7 mSv per year in the UK, but they are not zero, and cumulative exposure across multiple scans must be considered.
Because of this risk, ionising radiation techniques follow the ALARA principle — doses should be kept As Low As Reasonably Achievable. Particular caution applies to children (whose dividing cells are more radiosensitive) and to pregnant patients (where the foetus is especially vulnerable). Radionuclide imaging adds a further consideration: the patient continues to emit radiation after the scan until the tracer has decayed or been excreted, requiring temporary precautions around vulnerable individuals.
MRI carries no ionising radiation risk. The strong magnetic field and radiofrequency pulses used are not known to cause biological harm at diagnostic field strengths. However, the magnetic field creates a different safety concern: ferromagnetic objects become dangerous projectiles near the scanner, and implanted devices such as pacemakers or cochlear implants may malfunction or heat up. Careful screening of patients before entering the MRI suite is therefore essential.
Ultrasound is generally considered the safest modality. It uses non-ionising sound waves, carries no known biological risk at diagnostic intensities, and can be used throughout pregnancy. This is why ultrasound is the standard technique for foetal imaging — the safety profile is well established over decades of clinical use.
6. Putting It Together — Structured Comparison
Now that you have a clear understanding of each dimension, the skill the exam tests is combining them into a coherent, justified comparison. The table below gives you a reference framework. Use it to structure your answers, but always adapt your reasoning to the specific clinical context in the question.
Technique
Spatial resolution
Convenience
Ionising radiation?
Key safety concern
Plain X-ray
High (bone detail excellent)
Very high — fast, widely available
Yes (low dose)
Cumulative dose; avoid in pregnancy
CT
Very high (sub-mm slices)
High — fast but fixed equipment
Yes (moderate–high dose)
Higher dose than plain X-ray; avoid in pregnancy if possible
MRI
Very high (excellent soft tissue)
Low — long scan, restricted access
No
Ferromagnetic implants; claustrophobia
Ultrasound
Moderate (frequency-dependent)
Very high — portable, real-time
No
Operator dependence; limited penetration
Radionuclide / PET
Low (functional detail, not anatomical)
Low — tracer uptake wait; specialist centre
Yes (moderate dose)
Patient emits radiation post-scan; avoid in pregnancy
A common exam scenario presents a specific patient — perhaps a pregnant woman with abdominal pain, or a child with a suspected bone tumour — and asks you to justify a choice of imaging technique. The answer must address all three criteria: what resolution is needed for the diagnosis, how practical the scan is for this patient, and what the safety implications are given the patient's circumstances.
Worked example: Structuring a Comparison Answer
A 28-year-old pregnant patient (12 weeks) presents with right iliac fossa pain. The clinician considers ultrasound or CT to investigate possible appendicitis. Compare these two techniques in terms of resolution, convenience, and safety for this patient. [6 marks]
CT: very high spatial resolution, sub-millimetre slices, can clearly visualise the appendix and surrounding structures. Ultrasound: moderate resolution, dependent on operator skill and patient body habitus; the appendix may not always be visualised, particularly in obese patients or when obscured by bowel gas. — [B1] for stating CT has higher spatial resolution. [B1] for acknowledging ultrasound resolution is frequency-dependent and may be insufficient to visualise the appendix reliably.
Both are relatively convenient. CT is fast (minutes) and widely available. Ultrasound is portable, requires no special preparation, and can be performed at the bedside. Neither requires a long uptake period or specialist centre. — [B1] for a valid convenience comparison — accept any accurate point distinguishing the two (e.g. ultrasound is portable; CT requires a fixed scanner).
CT uses ionising radiation (approximately 10 mSv for abdominal CT). Ionising radiation poses a risk of DNA damage to the developing foetus, particularly in the first trimester when organogenesis is occurring. Ultrasound uses non-ionising sound waves and has no known harmful effects on the foetus at diagnostic intensities. — [B1] for identifying that CT uses ionising radiation and stating the risk to the foetus. [B1] for stating ultrasound is non-ionising and safe in pregnancy.
Ultrasound should be the first-line investigation. Although CT offers superior resolution, the ionising radiation risk to the foetus outweighs this advantage. Ultrasound is safe, convenient, and sufficient for initial assessment. CT may be used if ultrasound is inconclusive and the clinical need is urgent, as the risk of untreated appendicitis to both mother and foetus exceeds the radiation risk. — [B1] for a balanced conclusion that weighs safety against diagnostic need — this is the mark-winning evaluative judgement.
Answer: Ultrasound is preferred as first-line imaging due to its non-ionising nature and safety in pregnancy, despite lower resolution than CT. CT is reserved for cases where ultrasound is inconclusive and clinical urgency demands it.
7. Radionuclide Imaging — Resolution and Safety in Context
Radionuclide imaging deserves particular attention because its resolution and safety profile are both distinctive and frequently misunderstood. Recall from Section 3 that PET and gamma camera imaging have the lowest spatial resolution of the common modalities. This is not a flaw to be apologised for — it reflects the fundamental physics of detecting gamma photons emitted from within the body.
What radionuclide imaging provides instead is functional information: it shows where a particular biological process is occurring, not just where a structure is located. A PET scan using fluorodeoxyglucose (FDG) highlights regions of high glucose uptake, which correspond to metabolically active tumour tissue. This functional sensitivity can detect disease at a stage when the anatomy looks normal on CT or MRI — making radionuclide imaging complementary to, rather than competing with, high-resolution anatomical techniques.
In clinical practice, PET is frequently combined with CT in a single hybrid scanner (PET-CT). The CT component provides the anatomical map with high spatial resolution; the PET component overlays the functional signal. This combination addresses the resolution limitation of PET directly, and it is worth mentioning in exam answers when discussing how the limitations of radionuclide imaging are managed in practice.
On the safety side, radionuclide imaging involves administering a radioactive substance internally. The tracer emits ionising radiation from within the patient's body until it decays or is excreted. The effective dose varies with the tracer and the organ being imaged, but is typically in the range of 2–10 mSv — comparable to CT. Crucially, the patient remains a source of ionising radiation after leaving the scanner, which has implications for contact with pregnant women and young children in the hours following the scan.
8. Exam Technique — How to Score Full Marks on Comparison Questions
Comparison questions are among the most predictable in medical physics, but they are also among the most poorly answered. The most common reason for losing marks is addressing only one or two of the three criteria (resolution, convenience, safety) rather than all three. Examiners award marks for each criterion addressed, so a complete answer must cover all three explicitly.
A second common error is making vague comparative statements without scientific justification. Writing "MRI is safer than X-ray" earns no marks on its own. Writing "MRI does not use ionising radiation, so there is no risk of DNA damage, whereas X-ray delivers a dose of approximately 0.02 mSv which carries a small but non-zero risk of inducing cancer" earns the mark. Precision and mechanism are what distinguish a B-grade answer from an A-grade one.
For evaluate or justify questions, you must reach a conclusion. State which technique is more appropriate for the given scenario and explain why the advantages outweigh the disadvantages in that specific context. A conclusion that restates the comparison without making a judgement will not access the top mark band.
Sources and content standards
Published by Revise Right. Use the official specification reference to check the requirements for your course. Content updated: 2026-09-12.
A PET scan shows function rather than just anatomy. Using fluorodeoxyglucose, FDG, it highlights regions of high glucose uptake, which correspond to metabolically active tumour tissue, and can detect disease when the anatomy still looks normal on CT or MRI.
What is the radiation dose of a chest X-ray?
A chest X-ray delivers approximately 0.02 mSv, far less than natural background radiation of around 2.7 mSv per year in the UK. The dose is small but not zero, so cumulative exposure across multiple scans must still be considered.
Which medical imaging technique has the best resolution?
CT and MRI give the highest spatial resolution, with sub-millimetre slices in modern scanners; MRI also distinguishes between soft tissues far better. Ultrasound resolution depends on frequency: a 10 MHz probe resolves about 0.15 mm but cannot image deep organs. PET and gamma camera imaging have the lowest resolution, typically 4–10 mm for PET.
Which medical imaging techniques use ionising radiation?
X-ray, CT and radionuclide imaging, including PET, all use ionising radiation, which can damage DNA and carries a small cancer risk proportional to dose. Typical doses are about 0.02 mSv for a chest X-ray, 10 mSv for an abdominal CT and 3–4 mSv for a bone scan. MRI and ultrasound are non-ionising.
What is the safest medical imaging technique?
Ultrasound is generally considered the safest: it uses non-ionising sound waves, has no known biological risk at diagnostic intensities and can be used throughout pregnancy. MRI is also non-ionising, but its strong magnetic field turns ferromagnetic objects into projectiles and can make pacemakers or cochlear implants malfunction, so patients must be screened first.
Why is ultrasound used instead of CT in pregnancy?
CT delivers about 10 mSv for an abdominal scan, and ionising radiation risks DNA damage to the developing foetus, especially in the first trimester. Ultrasound is non-ionising and safe in pregnancy, so it is the first-line investigation, for example for suspected appendicitis, despite lower resolution. CT is kept for when ultrasound is inconclusive and the need is urgent.
Why is PET combined with CT?
PET has low spatial resolution but gives functional information: an FDG scan highlights high glucose uptake in metabolically active tumour tissue, detecting disease when the anatomy still looks normal. A hybrid PET-CT scanner overlays this functional signal on a high-resolution CT anatomical map, directly addressing PET's resolution limitation.
What is the resolution of a PET scan?
PET has a spatial resolution of typically 4–10 mm, and planar scintigraphy is coarser still, the lowest of the common modalities. The image is built from gamma photons emitted in all directions, and the statistical nature of radioactive decay limits localisation.
How long does an MRI scan take?
An MRI scan typically takes 20 to 60 minutes, and the patient must stay completely still inside a loud, narrow magnet bore. Claustrophobic patients may need sedation, which makes MRI the least convenient of the anatomical techniques.
How long does a CT scan take?
A full-body CT scan can be completed in under 10 minutes with modern multi-slice scanners, with no special preparation beyond removing metal objects. The patient must lie still inside a narrow bore for several minutes.
Who cannot have an MRI scan?
Patients with certain metallic implants, such as pacemakers or some surgical clips, cannot safely enter an MRI scanner, and implants such as cochlear implants may malfunction or heat up. Ferromagnetic objects become dangerous projectiles, so patients are carefully screened first.
What is the radiation dose of a bone scan?
A radionuclide bone scan delivers about 3–4 mSv, compared with about 0.02 mSv for a chest X-ray and around 10 mSv for an abdominal CT. Natural background radiation in the UK is around 2.7 mSv per year.
Why is an X-ray the first choice for fractures?
Plain X-ray is the first-line investigation for suspected fractures because it is the most convenient modality: widely available, fast, needing no preparation, with excellent bone detail at a low dose. Its limited soft-tissue contrast matters little for bone.
Is a patient radioactive after a nuclear medicine scan?
After radionuclide imaging, the patient continues to emit ionising radiation until the tracer has decayed or been excreted. Contact with pregnant women and young children needs care in the hours following the scan.
What is the difference between functional and anatomical imaging?
Anatomical imaging, such as X-ray, CT and MRI, shows where structures are, with high spatial resolution. Functional imaging, such as PET, shows where a biological process is happening, like glucose metabolism, sacrificing resolution for information anatomy alone cannot give.